Declaration Of Compatibility

Grounding Pad conformity

Download

SWAROPLATE Grounding Pads (Diathermy Pads / Neutral Electrodes) with preattached cables

Pre-attached grounding pads are pre connected with a cable for the connection between the HF Surgery Device and the neutral electrode. Simply select the right type of cable for your device for instant application.
Cardiology Chart Paper
The following types of SWAROPLATE Neutral Electrodes with preattached cables are available:

Item # Pad Use Size (length / width) Total Adhesive Surface Conductive Surface Impedance @ 600kHz approx. Temperature rise @ 700mA/60sec Base Electrode Connector Spec Sheets
6000839N Adults 215 x 104 mm 207 cm2 134 cm2 1 Ohm < 3°C 1201H Valleylab REM Spec Sheets
6000843N Adults 215 x 104 mm 207 cm2 134 cm2 1 Ohm < 3°C 1201H Jack Connector Spec Sheets
6000837N Adults & Children 156 x 126mm 144 cm2 105 cm2 1.5 Ohm   1210H Valleylab NON-REM Spec Sheets
6000841N Adults & Children 156 x 126mm 144 cm2 105 cm2 1.5 Ohm   1210H Jack Connector Spec Sheets
6000811N Adults 201 x 106 mm 200 cm2 144 cm2 1 Ohm < 3°C 1211 Valleylab REM Spec Sheets
6000835N Adults 206 x 122 mm 220 cm2 160 cm2 1 Ohm < 3°C 1212 Valleylab REM Spec Sheets
6000833N Children 154 x 88 mm 124 cm2 71 cm2 2 Ohm < 5°C 1222 Valleylab REM Spec Sheets
6000836N Children 164 x 124 mm 105 cm2 71 cm2 2 Ohm < 5°C 1226 Valleylab REM Spec Sheets
6000834N Infants and Prematures 165 x 74 mm 66 cm2

32 cm2

4 Ohm < 5°C 1223 Valleylab REM Spec Sheets
6000838N Adults & Children 160 x 122 mm 148 cm2 105 cm2 1,5 Ohm < 4°C 1220H Valleylab REM Spec Sheets
6000842N Adults & Children 160 x 122 mm 148 cm2 105 cm2 1,5 Ohm < 4°C 1220H Jack Connector Spec Sheets

Download our Grounding Pad Compatibility information here.


If you need assistance, do not hesitate to contact our knowledgeable customer service department at +44 1803 860100, or contact your local Sales Rep.

Terms of Use | Privacy Policy | Conditions of Sale | Site Map | Return Policy | Code of Ethics

Copyright Nissha Medical Technologies. All rights Reserved.

The OEM trademarks identified herein are the trademarks of the respective OEMs, and not Nissha Medical Technologies. Nissha Medical Technologies disclaims any affiliation, connection or association between it's products and those of the respective OEMs.